The differences on the energy response comply with the standards prescriptions. This same standard also sets requirements on the differences of the angular response R(α) in H p(10, α) for angles between 0° and ±75°: the variation does not exceed ±30% of the reference value. Figure 3.55b represents R(α) = H p(10, α)mes/H p(10, α)ref for angles between 0 and ±75° on the horizontal and vertical planes passing through the reference center of the detector. The horizontal angular Saphydose-n responses comply with IEC 1323 (±30%). Only the point + 60° from the vertical angular response shows a gap of 34%.

Add that the response of the dosimeter with respect to integrated personal equivalent dose for photonic radiation fields is less than 5% (test carried out with 60Co, 137Cs, X 164 and 250 keV). Where appropriate, in mixed field, the nature of the transmitted pulse allows the discrimination of photons and neutrons by the means of a suitable electronic treatment.

Finally, note that dosimeters are appropriate to specific radiological contexts: we will include the activation detectors used for criticality accidents (e.g. SNAC 2), composed of different activatable materials (see Chap. 2, Sect. 2.​3.​8). this class is insensitive to γ, enabling to characterize to the energy spectrum of the fluence rate and neutron emission, by the mean of a matrix algorithm.



Download



Copyright Disclaimer:
This site does not store any files on its server. We only index and link to content provided by other sites. Please contact the content providers to delete copyright contents if any and email us, we'll remove relevant links or contents immediately.